Connector
Connect the power cable here. Located   o n the back of the laser.
AC In
Emergency Stop Button
The emergency stop button is located on the top of the laser cart. Push this button to make the laser stop 
firing immediately.
This button works by turning off the pump and the high voltage power supply (hvps).
Reset the Emergency Button
Once pressed, the emergency stop button stays down. 
To reset the button back to its original position, twist the button clockwise (follow the arrows on the button).
Once the emergency stop button is reset, the laser needs 20 minutes to warm up before it can fire again.
Castors and Handles
The laser cart is on castors for easy movement. 
To lock any castor, lower the lever located above it. 
WARNING:
The laser cart also features handles for pushing and pulling. 
Components
WARNING:
Do not modify the power plug or use an adapter. 
This could cause an electrical hazard. If you need to use a 
different plug, contact Technical Support. For contact 
information, refer to Contact and Legal Information on 
page 477.
The emergency stop button
Lock all castors whenever the laser cart is not being moved.