If using a solution instead of a wipe, apply the solution to a cloth and wipe with the cloth instead of
l
spraying the solution directly on the system.
Make sure no solution gets into the laser ports. Do not wipe any exposed ports or connectors.
l
To Clean and Disinfect the Laser
1. Using an approved cleaner, clean the system to remove any visible gel, debris, or fluids.
2. Verify that all gel, debris and animal fluids have been removed from the laser. If necessary, repeat
the previous step with a new wipe.
3. Using an approved disinfectant, disinfect the system by wiping all surfaces that may have been
contaminated.
4. Ensure the laser stays wet with the disinfectant for the recommended wet contact time. This may
require reapplication to achieve the required wet time.
5. Allow the laser to air dry in a clean, well ventilated space. Do not dry with a cloth.
Caution:
of time. Leaving pools of disinfectant could lead to rust buildup.
Cleaning and Disinfecting Accessories
All accessories must be cleaned and disinfected after each exam.
Before You Begin
(If applicable) Disconnect the accessory from any wires or cables.
l
(If applicable) Turn off the accessory.
l
Wear appropriate personal protective equipment (PPE) recommended by the cleaning product
l
manufacturer, such as eyewear and gloves.
Precautions
Ensure that all cleaning and disinfecting products are appropriate for your facility's use, and are
l
compliant with the List of Compatible Cleaners and Disinfectants on page 446.
Do not use strong solvents such as thinner or benzene, or abrasive cleansers, since these will
l
damage the exterior surfaces.
Ensure that all cleaning and disinfecting products are not expired, and are in the right concentrations.
l
Always follow the manufacturer recommendations and local regulations when preparing, using, and
l
disposing of cleaning and disinfecting products.
Wipe from clean areas to soiled areas to avoid cross-contamination.
l
If using a solution instead of a wipe, apply the solution to a cloth and wipe with the cloth instead of
l
spraying the solution directly on the system.
Make sure no solution gets into any connectors or ports. Do not wipe any exposed ports or
l
connectors.
Some cleaners and disinfectants can cause an allergic reaction to certain individuals. Ensure that
l
the person handling the product is not allergic to them.
Maintenance
Do not allow disinfectant to pool on the laser for extended periods
Vevo F2 Imaging System User Manual