3. When prompted, select a desired location for the export and then tap Export.
The machine ID is then saved to a text file that can be sent to Technical Support.
To Update Your System License
Once you have received a new license file:
1. Save the license file on a USB storage device.
2. Open the System Information menu, and navigate to the System Information tab.
3. Tap Update License.
4. When prompted, enter administrator credentials.
5. In the file browsing panel on the left, navigate to the appropriate location for the USB storage device
and select the update file.
6. Tap Update. An Update Progress dialog appears.
7. Once the update is complete, tap Done.
The system will restart.
Cleaning and Disinfecting
The system, laser, transducer and a ccessories must be cleaned and disinfected after each imaging
session. It is important to follow these cleaning and disinfecting instructions without skipping any steps.
Note:
Please be aware that you must perform both a cleaning as well as a
disinfection for each component.