CAT (Cable and Antenna Test) Mode
CAT Mode Settings
This setting can be changed at any time without affecting calibration accuracy.
Averaging is often used to increase the dynamic range of a measurement.
To achieve the highest dynamic range, select NA mode and reduce the IF
Bandwidth setting. Learn more about dynamic range in
Range" on page 122

How to set Trace Averaging

1. Press BW 2.
2. Then
3. Enter a value using the numeric keypad. Enter 1 for NO averaging.
4. Press Enter.
5. While Trace Averaging is in process, press Sweep 3 then

Smoothing

Trace smoothing averages a number of adjacent data points to smooth the
peak-to-peak noise values on a displayed trace. The number of adjacent data
points that are averaged is known as the smoothing aperture. Aperture is set
by specifying a percentage of the X-axis span.
Trace smoothing does NOT significantly increase measurement time.
Smoothing can be used with any CAT format.
When enabled,
How to set Smoothing
1. Press BW 2.
2. Then
3. Then
4. Press Enter.

Single or Continuous Measure

This setting determines whether the FieldFox sweeps continuously or only once
each time the
or to allow you to save or analyze a specific measurement trace.
This setting can be changed at any time without affecting calibration accuracy.

How to set Single or Continuous

— Press Sweep 3.
Keysight N9927-90020 User's Guide
.
Average N
where N is the current count setting.
the averaging at 1.
Smo
appears on the FieldFox screen.
Smoothing ON
OFF.
Sm. Aperture
and enter a value between 0 and 25 (percent) using
the numeric keypad.
Single
button is pressed. Use Single to conserve battery power
"Increase Dynamic
Restart
to restart
65