Instrument Messages
Advisory Messages
Message
Band Adjust has no effect
on a Fixed marker
Band Adjust has no effect
with Mkr Function Off
Band-pass filter set to OFF
Cal Cancelled; Calibration
data cleared
Cal Invalid: meas freq
pt(s) > 3.6GHz are > 50MHz
from existing Cal pts
Carrier power is too low
for optimum dynamic range.
Connecting to source...
Demod Time is not available
in Zero Span
Detector <X> changed due to
physical constraints
Dynamic range is not
optimum. Set AUTO RF input.
Exp. Averaging not
available when AUTO PhNoise
is active.
FFT Width is not settable
unless Sweep Type is set to
FFT
File <filename> saved
Filter BW function is only
available for Gaussian
filter type
Instrument Messages
Description/Correction Information
If a Marker is a Fixed type marker, the marker's value does not change from when it
first became fixed. So you cannot change the band of a fixed marker.
If Marker Function is off changing the band has no effect
Turning on any high-pass or low-pass filter will turn off band pass filters.
User has canceled the cal either directly or indirectly by changing the setup
parameters. The current cal data has been erased. Perform a new user cal to obtain
calibrated results again.
When freq points being measured are above 3.6 GHz and a calibration has been
successfully performed, and the number of points are changed, the new points are
required to be within 50 MHz of the current cal points or the preselector optimize
frequencies become inaccurate and the whole cal needs to be invalidated.
Interpolation of the cal can only be performed if the new freq points are within 50
MHz of the cal points. To overcome this problem, change the number of freq points
back to match cal points or perform another user cal.
For better dynamic range, transmit band spur measurements require >10 dBm
signal power at the RF input port.
External Signal Generator is being sent SCPI commands interrogating it to see if it
is suitable for the MXA to control. Please wait until complete before pressing any
buttons.
The Demod Time function is not available in zero span because in zero span we are
ALWAYS demodulating.
You have selected more detectors than the instrument hardware can implement.
An existing detector selection has been changed to allow the current detector
choice to be selected.
<X> indicates the trace number for which the detector was changed.
You must select the FFT sweep type before you can set the FFT Width
The file save operation executed successfully.
Flattop and CISPR/MIL filters have defined shapes that cannot be altered. So only
the Gaussian filter type allows filter bandwidth definition changes.
37