SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
Figure 9-17 Display Line with Annotations
How to create and move a Display Line
1. Press Limit 8 >
2. Then enter a Y-axis value using the
Learn more,

Noise Marker

For comparison purposes, electronic noise measurements are often displayed
as though the measurement was made in a 1 Hz Res BW. However, making an
actual measurement at a 1 Hz Res BW is extremely slow.
A Noise Marker, unique to SA Mode, mathematically calculates the noise
measurement as though it were made using a 1 Hz bandwidth.
Several data points (or 'buckets') are averaged together to calculate the Noise
Marker readout. To accurately measure noise, the Noise Marker should NOT be
placed on, or too close to, a signal. The distance from a signal depends on
several factors. To know if an accurate reading is being made, move the Noise
Marker until consistent measurements are displayed in adjacent data points.
In addition, when a Noise Marker is displayed, the Detection method is
automatically switched to Average and
This occurs only when Detector is set to Auto. Learn more in
Method" on page
Keysight N9938-90006 User's Guide
Display Line OFF ON
press Enter. Or enter a value using the numeric keypad and press a suffix
key or press Enter.
"All about Limit Lines" on page
211.
arrows or the rotary knob, then
741.
PAvg
is shown on the FieldFox screen.
"Detection
213