Adjusting The Ceph Primary Diaphragm - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

4.5 Adjusting the cephalometer (XG
Preparations
4 – 60
Plus
ready
/ 3D
/ XG 5, if ceph present)
4.5.3

Adjusting the ceph primary diaphragm

Menu: Diaphragm/system adjustment
i
NOTE
For units with a unit software < V02.27 the exposure parameters for the prima-
ry diaphragm adjustment may deviate from the parameters shown here
(64 kV / 16 mA; 6.1 s).
CAUTION
It is essential that the PAN needle phantom is removed from the bite block
holder of the panoramic X-ray unit before a Ceph exposure is taken; otherwise
the phantom may collide with the sensor.
Move the ear plug holders on the cephalometer completely apart and
swing them out of the beam direction (ap).
Insert the test phantom in the sensor slot on the panoramic X-ray unit
(see page 4-59).
Plug the sensor into the sensor slot on the cephalometer.
/
Open the
DIAPHRAGM
(see page 4-5).
Ceph - Primary diaphragm
menu
SYSTEM ADJUSTMENT
D3352.076.01.26.02 09.2016
Tab 4
59 38 399 D3352