Service Manual GALILEOS
Chassis:
X-ray tube assembly:
Detector:
Facescanner (optional):
Geometry:
Scanning operation:
61 25 665 D3437
D3437.076.01.15.02 06.2012
3.2.7

Technical data

Technical data
Model designation
Nominal voltage:
Permissible fluctuation:
Permissible drop under load:
Rated current:
Nominal power output:
Current time product:
Nominal frequency:
Internal line impedance:
Main building fuse:
Power consumption:
Focal spot size acc. to IEC 60336,
measured in the central X-ray beam:
kV:
mA:
Pulsed mode:
Total filtration of X-ray tube assembly
Cone-beam angle:
High voltage generation frequency:
Type: Image intensifier (I.I.),
Thales or Siemens
Active input window size:
Camera:
Model designation
Maximum current:
Power consumption:
Weight
Source-I.I. converter coating distance
(central X-ray beam)
Source-isocenter distance
(central X-ray beam)
Source-skin distance
(minimum distance)
Orbital angle
GALILEOS
200 V – 240 V
±10%
10%
6 A
0.6 kW at 85 kV/7mA
42 mAs
50 Hz/60 Hz
max. 0.8 ohms
25 A slow-blow (16 A for single line)
0.9 kVA
0.5
85 kV
5 mA/7 mA
10 ms – 30 ms
> 2.5 Al / 90 IEC 60522
collimated to approx. 24°
80 kHz – 100 kHz
215 mm (8 1/2") diameter
2
Pixels: 1000
FPS: 15 – 30
Dynamics: 12 bits,
(4096 brightness values), 60 dB
Facescan
6.25 A
0.945 kVA
5.7 kg
510 mm (20 1/16")
333 mm (13 1/8")
approx. 220 mm (8 5/8")
204°
3 Unit description
3.2 Hardware
45