Technical Data - Sirona Orthophos XG 5 Operating Instructions Manual

Hide thumbs Also See for Orthophos XG 5:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG 5 / Ceph
60 04 811 D3352
D3352.201.04.25.23 12.2017
3.2

Technical data

Unit data
Model designation:
Nominal voltage:
Permissible fluctuation:
Permissible drop under load: 10%
Rated current:
Rated power:
Nominal frequency:
Mains resistance:
Main building fuse:
Power consumption:
Power output of tube
assembly:
Tube voltage:
Tube current:
Maximum setting range:
High-voltage waveform:
High voltage generation
frequency:
Program duration:
Exposure time:
Image acquisition scale:
Exposure time for a
cephalometric image:
Image acquisition scale for a
cephalometric image:
Total filtration of X-ray tube
assembly:
Focal spot size as specified
in IEC 60336, measured in
the central X-ray beam:
ORTHOPHOS XG 5 / XG 5 DS / Ceph
200 – 240 V
± 10%
12 A
2 kW at 90 kV/12 mA with any radiation
time
50 Hz / 60 Hz
max. 0.8 ohms
25 A slow-blow (16 A for single line)
2 kVA
90 kV/12 mA = 1080 W with any radiation
time
60 – 90 kV (at 90 kV max. 12 mA)
3 – 16 mA (at 16 mA max. 66 kV)
60 kV / 3 mA to 90 kV / 12 mA
High-frequency multipulse
Residual ripple ≤ 4 kV
40 – 120 kHz
See "Program values" [ → 93].
See "Program values" [ → 93].
For P1, normal dental arch (slice center)
approx. 1:1.19, i.e. the acquired image is
magnified by approx. 19% on average
compared to reality.
14.9 s max.
approx. 1:1.1, i.e. the acquired image is
magnified by approx. 10% on average
compared to reality.
> 2.5 AI / 90 IEC 60522
0.5 mm
3 Unit description
19

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos xg 5 ds

Table of Contents