7.3.8

Geometry calibration

Geometry calibration
IMPORTANT
For an installed and configured Facescan:
The face scanner is automatically calibrated during the geometry
calibration of GALILEOS. In this case, make sure that the normal room
lighting is switched on during the calibration process. The room does not
have to be darkened during calibration.
"Service functions"
The
"Geometry"
1. Click the
The corresponding tab card is selected.
2. Read off the serial number of the geometric phantom from the ID
label of the phantom and enter it in the
text box of the
3. Insert the geometric phantom in the bite block holder of the
unit [ → 164].
4. For Facescan: Check to make sure that the room lighting is switched
on. Switch it on if necessary.
5. Click on the
SIDEXIS XG makes the unit ready for exposure [ → 162].
Service routine S010.13 is displayed on the control panel.
6. Press the R key to move the unit back to the starting position.
7.3 Adjustment and calibration via the "Service Functions" menu
menu is called [ → 161].
tab.
"Geometry"
tab card.
"Image acquisition"
button.
7 Adjusting/calibrating the unit
"Seriennr. Kalbrierphantom"
187