4.5 Adjusting the cephalometer (XG
Preparations
Position of the ear plug holders
Fitting the adjusting caps
D
4 – 102
Plus
ready
/ 3D
/ XG 5, if ceph present)
4.5.7
Checking and adjusting the alignment of the
ear plugs
Checking the ear plug alignment
i
NOTE
The sensor must be plugged into the sensor slot on the cephalometer.
No sensor must be plugged into the slot on the panoramic X-ray unit.
Plus
ready
For XG
/ 3D
:
Select the
mode in the main menu on the Easypad.
CEPH
For XG 5:
Select one of the ceph programs on the Multipad.
1.
Move the ear plug holders completely apart and
swing them into the beam direction.
D
2.
Fit adjusting caps
onto the ear plugs and secure them with adhesive
tape.
Fit the black adjusting cap on the outside (sensor side)
Fit the transparent adjusting cap on the inside (tube assembly side).
Tab 4
59 38 399 D3352
D3352.076.01.26.02 09.2016