Adjusting The Panoramic X-Ray Unit - Sirona Orthophos XG 3 Installation Instructions Manual

Hide thumbs Also See for Orthophos XG 3:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Instructions ORTHOPHOS XG 3

11.2Adjusting the panoramic X-ray unit

1.
11.2.1Diaphragm/system adjustment menu
/
The DIAPHRAGM
SYSTEM ADJUSTMENT menu guides
you through the adjustment of the panoramic unit. This ser -
vice routine is started from SIDEXIS XG:
U
C
TILITIES
ONSTANCY TEST
X-
S
RAY DEVICE
ERVICE EXPOSURE
S
X-
INPUT
ELECT
RAY COMPONENT
D
IAPHRAGM TEST EXPOSURE
IMPORTANT
SYSTEM ADJUSTMENT menu is pass -
/
The DIAPHRAGM
word-protected. As password, enter the first four digits of
the current system date (PC) in reverse order.
Example: On 05/30/2004, the service password is 5003
IMPORTANT
-
The SELECT X
RAY DEVICE and SELECT X
NENT dialogs appear only if several x-ray devices or x-ray
components are available for selection.
60 51 416 D 3352
D 3352.031.03.18.02
S
XCXP
ELECT
P
ASSWORD
-
-
RAY COMPO
11.2 Adjusting the panoramic X-ray unit
2.
S E R V I C E
IMPORTANT
When you open the DIAPHRAGM
menu, the unit switches from the user mode to the PC ser -
vice mode logged by the PC. The service mode is indicated
by the SERVICE display on the Multipad (2.).
In the PC service mode, the control options that are avail -
able on the Multipad are determined by SIDEXIS and the
currently selected service routine. General control of the
unit by means of the Multipad (as in the user mode) is not
possible in this mode.
/
The DIAPHRAGM
SYSTEM ADJUSTMENT menu has
5 submenus:
PAN - Sensor adjustment
PAN - Diaphragm
PAN - Symmetry
PAN - Reset adjustment
Sirona Service Sheet (for internal use only)
You can change between the individual submenus by click -
ing the menu tabs with the mouse. To quit the DIA
/
PHRAGM
SYSTEM ADJUSTMENT menu, click CANCEL .
11 Appendix
/
SYSTEM ADJUSTMENT
-
109