Example Of Error Logging Data - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

Tab 3
System time
Entry type
Entry data [Message]
Entry data [Error]
Entry data [Error Sidexis]
Entry data [Stringname]
Entry data [Stringsegment]
59 38 399 D3352
D3352.076.01.26.02 09.2016
3.1.1

Example of error logging data

2006-03-06, 20:13:02
[Message]
[Error]
[Error Sidexis]
[Stringname]
[Stringsegment]
Recording started
Value: 9000
Recording stopped
Recording canceled
Logbook started
Image state switched to Released
Other entry data which document the occurrence of a rescue event include:
– Image state switched to Rescue
– Rescue request Sidexis Error
– Rescue request Sidexis TrackEpilogue
– Rescue request Sidexis Timeout
These entry data may also occur after "Recording stopped" or "Cancel" and
indicate an exceptional circumstance. You can supply important information
for error diagnostics in coordination with the Sirona Customer Service Center.
E6 07 06
ERR_DX7_TTP_LOST
SidErr: ERR_SOCKET_ERROR
SockErr:
Key Act
Key OK
Counter (XG 3 PPE)
7YFWDUFV-
E4MMRJBW
061-00133
3.1 Error logging memory
:
System time (clock on the
DX11)
:
General system event
:
Error event
:
Network error event
:
Free status texts
:
Additional data (string names)
:
Beginning of a recording
:
Sequence ID of recording
:
End of a recording
:
Recording cancellation
:
Corresponds to power-on of
Unit
:
Recording has been delivered to
and confirmed by SIDEXIS
:
Error code
:
Clear text display of error
:
Details on
network error
(for Sirona only)
:
Details on
network error
(for Sirona only)
:
Activation transaction
:
Activation transaction
:
Current counter reading
:
e.g. Activation/
confirmation code
(for activation transaction)
:
Example: Counter
(ID counter reading)
3 – 5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents