8.3

Troubleshooting of simple defects

In case of malfunctioning, proceed as follows:
First general directions in case of malfunctioning:
● Check if power supply is connected properly and/or rechargeable
battery is loaded.
● Check if fiber is installed correctly.
● Be sure that all operational steps have been carried out correctly.
● Check the functioning of the finger switch by pressing several times.
The touch screen of the SIROLaser Xtend remains dark after switching
it on.
● Check the connection of the power cable and/or check the
rechargeable battery.
● Make sure that the switch on the switching power supply is switched
on.
● Check the connection of the interlock device.
The finger switch is claimed to be missing.
● Check if the plug of the cable is properly connected to the control
unit.
The fiber is claimed to be missing.
● Make a visual check of the optical fiber and its connector. If you see
any damage (e.g. scratches) replace the optical fiber with a new
one.
● Check the connection of the optical fiber.
● Check the proper assembly of the handpiece sleeve.
● Be sure that all operational steps have been carried out correctly.
There is no aiming beam.
● Check to see if the optical fiber or its connector is damaged. If the
optical fiber is damaged, replace it with a new one.
● Check the connection of the optical fiber.
● Check the proper assembly of the handpiece sleeve.
● Be sure that all operational steps have been carried out correctly.
The light of the aiming beam does not project a uniform circular pattern.
● Trim the end of the optical fiber with the fiber cutter again. Always
make the notch perpendicular to the optical fiber.
8 Maintenance and service
8.3 Troubleshooting of simple defects