Troubleshooting Of Simple Defects; D3648 - Sirona sirolaser blue Operating Instruction

Laser
Hide thumbs Also See for sirolaser blue:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Operating Instructions SIROLaser Blue
General
Finger switch
Foot control
Single-use fiber tip or therapy light guide

65 40 863 D3648

D3648.201.01.02.02 10.2015
NOTICE
If national or local legal regulations require additional safety checks for
your laser unit, these regulations must be complied with and the
corresponding checks must be performed.
The manufacturer accepts responsibility for the safety of the laser unit
only if the following requirements are fulfilled:
● Modifications of the laser unit or repair work may be performed only
by authorized personnel.
● The electrical installations in the rooms where the SIROLaser Blue is
used must fulfill the applicable legal requirements.
● The unit must be used in compliance with the instructions provided in
the present manual.
8.3

Troubleshooting of simple defects

In case of malfunctioning, proceed as follows:
First general directions in case of malfunctioning:
● Check if power supply is connected properly and/or rechargeable
With foot control
battery is loaded.
● Check if EasyTip or MultiTip is installed correctly.
● Be sure that all operational steps have been carried out correctly.
● Check the functioning of the finger switch and/or wireless foot control
by pressing several times.
The touch screen of the SIROLaser Blue remains dark after switching it
on.
● Check the connection of the power cable and/or check the
With battery
rechargeable battery.
● Make sure that the switch on the switching power supply is switched
on.
● Check the connection of the interlock device.
The finger switch is claimed to be broken.
● Check if the finger switch is chosen in the settings submenu.
With foot control
● Check if the plug of the cable is properly connected to the control unit.
The foot control is not working or claimed to be missing.
With foot control
● Check if the foot control is chosen in the settings submenu.
● Check the battery of the wireless foot control.
● Register the wireless foot control again.
The single-use fiber tip or therapy light guide is claimed to be missing.
● Make a visual check of the single-use fiber tip or therapy light guide
and its connector. If you see any damage (e.g. scratches) replace the
single-use fiber tip or therapy light guide with a new one.
● Check the connection of the single-use fiber tip or therapy light guide.
● Check the proper assembly of the handpiece sleeve.
8 Maintenance and service
8.3 Troubleshooting of simple defects
65