Agilent Technologies E5071C Manual page 1048

Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Programming
170 OUTPUT @Agte507x;":SYST:PRES"
180 OUTPUT @Agte507x;":SENS1:FREQ:STOP ";Stop_freq
190 OUTPUT @Agte507x;":SENS1:SWE:POIN ";Nop
200 !
210 OUTPUT @Agte507x;":CALC1:TRAN:TIME:LPFR"
220 !
230 OUTPUT @Agte507x;":CALC1:PAR1:DEF "&Para$
240 OUTPUT @Agte507x;":TRIG:SOUR BUS"
250 !
260 ! 1 Port Full Calibration (ECal)
270 !
280 PRINT "Connect Port 1 to ECal Module. Then push [Enter] key."
290 INPUT "",Buff$
300 OUTPUT @Agte507x;":SENS1:CORR:COLL:ECAL:SOLT1 1"
310 OUTPUT @Agte507x;":SYST:ERR?"
320 ENTER @Agte507x;Buff$
330 !
340 ! Measurement
350 !
360 PRINT "Set DUT. Then Push [Enter] key."
370 INPUT "",Inp_char$
380 !
390 OUTPUT @Agte507x;":TRIG:SING"
400 OUTPUT @Agte507x;"*OPC?"
410 ENTER @Agte507x;Buff$
420 !
430 OUTPUT @Agte507x;":DISP:WIND1:TRAC1:Y:AUTO"
440 PRINT "Push [Enter] key. -> [Time Domain Transform]"
450 INPUT "",Inp_char$
460 !
470 ! Time Domain Transform
480 !
490 OUTPUT @Agte507x;":CALC1:TRAN:TIME "&Tran_type$
500 OUTPUT @Agte507x;":CALC1:TRAN:TIME:STIM "&Stim_type$
510 OUTPUT @Agte507x;":CALC1:TRAN:TIME:KBES ";Win_beta
520 OUTPUT @Agte507x;":CALC1:TRAN:TIME:STAR ";Star_time
530 OUTPUT @Agte507x;":CALC1:TRAN:TIME:STOP ";Stop_time
1275