Status[N]:Operation:enable; Status:preset; Status:questionable[:Event] - Keysight N774 C Series Programming Manual

Multiport power meter
Table of Contents

Advertisement

Command:

:STATus[n]:OPERation:ENABle

Syntax:
:STATus[n]:OPERation:ENABle<wsp><value>
Description:
Sets the bits in the Operation Slot Status Enable Mask (OSSEM) for slot n that enable the contents of the Operation Slot
Status Event Register (OSSER) for slot n to affect the OSESR.
Setting a bit in this register to 1 enables the corresponding bit in the OSSER for slot n to affect bit n of the OSESR.
Parameters:
The bit value for the OSSEM as a 16-bit unsigned integer value (0 .. +65535)
Response:
none
Example:
:stat0:oper:enab 128
Command:
:STATus[n]:OPERation:ENABle?
Syntax:
:STATus[n]:OPERation:ENABle?
Description:
Returns the OSSEM of slot n
Parameters:
none
Response:
The bit value for the OSSEM as a 16-bit unsigned integer value (0 .. +65535)
Example:
:stat0:oper:enab? -> +128
Command:

:STATus:PRESet

Syntax:
:STATus:PRESet
Description:
Presets all bits in all OPERation and QUEStionable status systems to 0.
Parameters:
none
Response:
none
Example:
:stat:pres
Command:

:STATus:QUEStionable[:EVENt]?

Syntax:
:STATus:QUEStionable[:EVENt]?
Description:
Returns the Questionable Status Event Summary Register (QSESR).