1
Specifications and Characteristics
Table 1-7
Number of averages
Settling time (s)
(Normal mode)
Settling time (s)
(x2 mode)
[a] E-Series power sensors in Fast mode (using free run trigger), within the range –50 dBm to +17 dBm, the settling
Figure 1-2
36
E441x and E9300 Series settling time
1
0.08
0.07
time is:
N1913B: 10 ms
N1914B: 20 ms
Normal
mode
mode
80 ms
Typical
130 ms
settling
times
1.1 s
450 ms
6.7 s
E441x Series settling time with auto-filter
Keysight N1913B/N1914B EPM Series Power Meters Service Guide
2
4
8
0.13
0.24
0.45
0.09
0.15
0.24
x2
70 ms
+50 dB
90 ms
+10 dB
+10 dB
3.5 s
+10 dB
[a]
16
32
64
128
1.1
1.9
3.5
6.7
0.45
1.1
1.9
3.5
Max dBm
Sensor
dynamic
range
Min dBm
256
512
1024
14
27
57
6.7
14
27