Initiate:continuous; Initiate[:Immediate]; List:band[1-201] - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
Open "C:\TEST.gif" For Binary As #1
For i = 0 To Nop - 1
Next i
Close
Equivalent key
No equivalent key is available on the front panel.
Pressing [Save/Recall] - SAVE DISPLAY key saves the current screen
image into the USB memory.

:INITiate:CONTinuous

Syntax
:INITiate:CONTinuous {ON|OFF|1|0}
:INITiate:CONTinuous?
Description
Enables the automatic trigger to change state from the "Idle" state to the "Wait
for Trigger" state. Refer to the chapter on Remote Control.
Parameter
ON or 1
OFF or 0 (Preset value)
Equivalent key
No equivalent key is available on the front panel.

:INITiate[:IMMediate]

Syntax
:INITiate[:IMMediate]
Description
Initiates the trigger to change from the Idle" state to the "Wait for Trigger"
state one time. Refer to the chapter on Remote Control. (No Query)
Equivalent key
No equivalent key is available on the front panel.

:LIST:BAND[1-201]

Syntax
:LIST:BAND[1-201] {A|B|OFF},<low>,<high>
:LIST:BAND[1-201]?
352
Put #1, , GifData(i)
Description
Enables automatic trigger state change
Disables automatic trigger state change
Keysight E4980A/AL Precision LCR Meter

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4980aalE4980al

Table of Contents