Initiate[:Immediate]; List:band[1-201] - Keysight E4980AL User Manual

Precision lcr meter
Hide thumbs Also See for E4980AL:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
OFF or 0 (Preset value)
Equivalent key
No equivalent key is available on the front panel.

:INITiate[:IMMediate]

Syntax
:INITiate[:IMMediate]
Description
Initiates the trigger to change from the Idle" state to the "Wait for Trigger"
state one time. Refer to the chapter on Remote Control. (No Query)
Equivalent key
No equivalent key is available on the front panel.

:LIST:BAND[1-201]

Syntax
:LIST:BAND[1-201] {A|B|OFF},<low>,<high>
:LIST:BAND[1-201]?
Description
Sets the limit values of the limit function for the list sweep measurement. When
A or B is the parameter, both lower and the upper limit should be input. If
either of them is not input, error No. 109 is returned. In the case of
OFF,<low>,<high> are optional.
Parameter
A
B
OFF (Preset value)
Preset value
Unit
Equivalent key
[Meas Setup] - LIST SETUP - LMT - A
[Meas Setup] - LIST SETUP - LMT - B
[Meas Setup] - LIST SETUP - LMT - 
Keysight E4980A/AL Precision LCR Meter
Description
Disables automatic trigger state change
Description
Sets limit test target to A (Primary parameter)
Sets limit test target to B (Secondary parameter)
Sets limit test target to OFF
<low>,<high>
Query returns -9.9E37, +9.9E37 for <low>, <high>,
respectively in the case of OFF.
Depends on
:FUNCtion:IMPedance[:TYPE]
357