[:SENSe] Subsystem

The [:SENSe] subsystem commands are divided into several sections. Each
section, or subtree, deals with controls that directly affect
instrument-specific settings and not those related to the signal-oriented
characteristics.
[:SENSe]:AVERage[STATe] . . . <Boolean>
COMMAND
Turns averaging ON and OFF. When averaging is ON, each new valid
measurement result is the average of the number of measurements specified in
the AVERage:COUNt command.
– An ABORt command interrupts the averaging and prevents a valid
COMMENTS
– Averaging cannot be turned ON when the value of AVERage:COUNt is one.
Avg
RELATED
FRONT-PANEL
KEYS
[:SENSe]:AVERage:COUNt . . . <numeric_value>
COMMAND
Specifies the number of measurements to combine when AVERage:STATe is ON.
The acceptable range for the <numeric_value> parameter is 1 to 99.
RANGE
When averaging is ON, some devices may automatically set :COUNt values in the
COMMENT
TRIGger subsystem based on the AVERage:COUNt value. This is done to ensure
that the TRIGger subsystem provides enough triggers for the average.
Avg
RELATED
FRONT-PANEL
KEYS
[:SENSe]:CORRection:CSET:SELect . . . <name>
COMMAND
Selects a power calibration table by name from eight available profiles.
Valid table names: 8481A, 8481D, 8482A, 8485A, 8487A, CUST1, CUST2, CUST3
– *RST: OFF
COMMENTS
232
measurement result.
Attempting to turn averaging ON when AVERage:COUNt=1 causes an error.