Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict; Signal
Track is not available
with Continuous Peak
-221
Settings conflict; Signal
Track is only available in
Swept SA measurement
-221
Settings conflict; Signal
Track is turned off when
Zero Span is selected
-221
Settings conflict; Source
List Step<n> MPA GPS port
power setting <n>dBm is
lower than -130dBm
minimum
-221
Settings conflict; Source
List Step<n> MPA TX
port<n> amplitude
correction value <n>dB is
out of range. The valid
range is <n> ~ <n>dB"
-221
Settings conflict; Source
List Step<n> MPA TX
port<n> amplitude
correction delta exceeds
<n>dB between port<n1>
and port<n2>
-221
Settings conflict; Source
List Step<n> MPA TX
port<n> power setting
<n>dBm is lower than
-130dBm minimum
-221
Settings conflict; Source
Output is not available
while Output Port is None.
-221
Settings conflict; Span
limited to XXX
-221
Settings conflict; Span
Zoom is not available in
Zero Span
-221
Settings conflict; Step
duration cannot be set to
Play Count when the step
is generating a CW tone
Instrument Messages
Verbose/Correction Information
The signal tracking feature cannot be used while you are also
using the continuous peak function.
The signal track functionality can be used when making a swept
SA measurement. It is not available in the SA measurement when
you are using FFT sweeps.
Signal Track is not available when you have selected Zero Span.
So if Zero Span is entered while in Signal Track is On, Signal Track
is turned off.
Indicating the power output setting (UI power + Amp Corr Value)
on MPA GPS port on Source List Step<n> is lower than
supportable minimum -130dBm.
Set power output setting to be larger than -130dBm to eliminate
this warning.
Indicating the amplitude correction value of MPA TX port<n> on
Source List Step<n> is out of range.
Set related amplitude correction value within the valid range
given in error message.
Indicating the amplitude correction delta value between MPA TX
port<n1> and port<n2> on Source List Step<n> (port difference)
exceeds max loss delta value.
Set related amplitude correction delta value to be lower than the
max loss delta value given in error message.
Indicating the power output setting (UI power + Amp Corr Value)
on MPA TX port<n> on Source List Step<n> is lower than
supportable minimum -130dBm.
Set power output setting to be larger than -130dBm to eliminate
this warning.
You are attempting to enable the Source Output power while no
Output Port is selected; first select the Output Port.
Span Zoom does not work with a time domain x-axis. You must
select a span greater then 0 Hz.
The you tried to set the source list step duration type to Play
Count while the waveform of the step is CW. Play Count only
applies to ARB.