S017: Test Step 23(Xg 3 Ppe Only) - Sirona ORTHOPHOS XG 3Dready Service Manual

Table of Contents

Advertisement

Configuration of PPE display warning level
5.13 Service routine S017
Defining the warning level for display of
remaining exposures after each expo-
sure
Quitting the service routine
5 – 120
5.13.18

S017: Test step 23(XG 3 PPE only)

Start test step 23 in the same way as described on page 5-85 or by press-
ing the Service key
i
NOTE
If you go to test step 23 from test step 22 by pressing the Service key, the se-
curity access confirmation will be skipped.
1.
Use the arrow keys in selection field 1 to select the number of remaining
exposures starting with which the warning (indicating the number of expo-
sures left) will be displayed after each exposure.
XG 3_PPE: Multipad
______30________
1
2
2.
To save your selection, first press the memory key
up) and then press the
XG 3_PPE: Multipad
______50________
1
2
R
To return to the Service menu, press the double arrow key
To quit the Service menu and return to the main menu, press the double
arrow key
.
in test step 22.
3
R key
.
R
3
D3352.076.01.26.02 09.2016
Tab 5
(the R key lights
.