PD1550A System Calibration- AutoCal
2 Low-Side Gate Voltage Source Calibration -- V
Calibrates voltage gain and offset for the Low-Side Gate Drive.
Hardware Setup:
- Remove the DUT module from the Interface Module before calibration.
- Install the Low-Side Gate Drive Module on the Interface Module.
- Install a low value resistance Low-Side Gate Resistance module (for example,
- Install UDI cables on the Low-Side Gate Drive Module and Interface Module.
- Use a red calibration cable to connect the SMU red Force terminal to either of
- Use a black calibration cable to connect the SMU black Force terminal to
Figure 10 Low-Side Gate Voltage Source Calibration Connections
Procedure:
108
2.2 ).
the VG pins on the Low-Side of the Interface Module.
either of the KS pins on the Low-Side of the Interface Module.
To Low-Side
KS terminal
Click Start to perform the calibration and continue to next step.
2 Low-Side Gate Voltage Source Calibration -- V
GS,L
To Low-Side
VG terminal
Keysight PD1550A Installation Guide
Output
GS,L
Output