S007: Error Logging Memory; S007: Test Step 1 - Sirona Orthophos XG 3D Ceph Service Manual

Table of Contents

Advertisement

8 Perform service routines via the control panel
8.7 S007: Error logging memory
SR*
S007
S007.1
S007.2
S007.5
280
8.7

S007: Error logging memory

SHZ**
Function
Error logging memory
No
Display error logging memory
Yes
Clearing error logging memory
No
Enabling CAN bus logging in the web interface
* SR=service routine, ** SHZ=security access
8.7.1

S007: Test step 1

Display error logging memory
In addition to service routine S007.1, you can also use the extended detail
query in SiXABCon to check the error logging memory.
Symbol on the control
panel
Patient symbol key 1 (B)
Patient symbol key 2 (C)
Patient symbol key 3 (D)
Selection field
1
2
3
* Factory setting
1. Call service routine S007.1.
2. Use the patient symbol keys (B, C, D) to select the step width for
scrolling between the error events (see table).
Service Manual ORTHOPHOS XG 3D / Ceph
Status
is selected
is selected
is selected
Selection/display
Error event
Error code for the selected event
Date and time of the selected
error event
The selected patient symbol key lights up.
Sirona Dental Systems GmbH
Function
Step width for scrolling
between error events = 1*
Step width for scrolling
between error events = 10
Step width for scrolling
between error events = 100
63 03 510 D3352
D3352.076.03.11.02 09.2016