Dimensions/Space Requirements; Front View; Sirona Dental Systems Gmbh - Sirona Orthophos XG 3D/Ceph Installation Manual

Table of Contents

Advertisement

Sirona Dental Systems GmbH

Installation Manual ORTHOPHOS XG 3D / Ceph
2249
88 1/2"
63 03 452 D3352
D3352.031.05.08.02 10.2015
3.2

Dimensions/Space requirements

3.2.1

Front view

NOTICE
Unit can travel to the ceiling.
The minimum ceiling height should be 2.10 m (82 3/4"). If the ceiling
height is lower than 2.27 m (89 1/2") (max. travel height of 2.25 m (88 1/
2")), the travel height of the unit must be adjusted or limited prior to
startup via service routine S018.2.
IMPORTANT
The measurements on the drawings are specified in both mm and
inches.
Front view: XG 3D
1950
76 3/4"
402,5
15 7/8"
A
Alternative fastening if it is not possible to screw the unit onto the
floor. Order bracket separately.
These dimensions apply to installation of the unit without the floor stand.
Installation with the floor stand results in an additional 30 mm (1 1/4")
increase of all height dimensions (see also separate Installation
Requirements document).
3.2 Dimensions/Space requirements
OR THOPHOS
A
3 Unit description
15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents