Function Specific Troubleshooting
Step 12. Use the softkeys to select the SCANNER INTERFACE test.
Softkey
INCR +
DECR -
Step 13. Press the EXECUTE - TEST START softkeys to execute the SCANNER
INTERFACE test.
Step 14. Confirm that LEDs on the scanner simulator board turn ON in accordance
with the E4980A/ E4980AL settings displayed in the LCD.
Step 15. Press the TEST STOP softkey.
Do not execute any SELF TEST except for the Scanner Interface Test or the
E4980A/ E4980AL will become inoperative.
Step 16. Return the switch settings on the scanner interface board to their original
settings.

To Check the Bias Current Interface Function (Option 002 only)

This test verifies the bias current interface functions.
Equipment
Table 4-10
Equipment List of the Bias Current Interface Function Test (Option 002 only)
Bias Interface Simulator
Bias Interface Cable
DC Power Supply
Adapter
Procedure
Step 1. Set all switches of S1 and S2 on the bias interface simulator to '1' as shown in
Figure
60
Function
Increments the test number of the Test Menu.
Decrements the test number of the Test Menu.
4-15.
Keysight p/n 42841-65001
Keysight p/n 42841-61640
Keysight E3610A
Keysight p/n E4980-60102