'Single sweep mode
status = viWrite(analyzer, "SYST:DISP:UPD OFF", 17, retCnt)
'OFF: display off
'--------- Perform measurement and query results ------------
cmd$ = "MPOW? "
cmd$ = cmd$ + "935.2 MHZ,"
'Center frequency
cmd$ = cmd$ + "1MHZ,"
'Resolution bandwidth
cmd$ = cmd$ + "434US,"
'Measurement time
cmd$ = cmd$ + "VID,"
'Trigger source
cmd$ = cmd$ + "50PCT,"
'Trigger threshold
cmd$ = cmd$ + "1US,"
'Trigger offset, must be > 125 ns
cmd$ = cmd$ + "PEAK,"
'Peak detector
cmd$ = cmd$ + "8"
'Number of bursts
status = viWrite(analyzer, cmd$, 47, retCnt)
status = viRead(analyzer, result$, 100, retCnt)
'Read results
Print result$
END SUB