To properly use this command, you have to specify the unit of the trigger level
(DBM or PCT).
<trigger offset>
Offset between the detection of the trigger signal and the start of the measurement.
The range is from 125 ns to 100s
<type of meas>
Determines whether mean power (RMS) or peak power (PEAK) is to be measured.
The detector is selected accordingly. Possible values are MEAN and PEAK.
<# of meas>
Number of individual bursts to be measured. The range is 1 to 625.
Return value:
The query command returns a list separated by commas (comma separated values
= CSV), which contains the power measurement results in floating-point format.
The unit used for the return values is always dBm.
The command
"SENSe:MPOWer? 935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20"
may, for instance, cause the following list to be returned:
18.3,18.6,18.1,18.0,17.9,18.3,18.6,18.1,18.0,17.9,18.3,18.6,18.1,18.0,17.9,18.3,
18.6,18.1,18.0,17.9
Example:
"SENSe:MPOWer 935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20"
'performs a measurement sequence with the following settings:
'Frequency = 935.2 MHz,
'Resolution bandwidth = 1 MHz
'Measurement time = 434 µs
'Trigger source = VIDEO
'Trigger threshold = 50%
'Trigger offset = 5 µs
'Type of measurement = MEAN power
'No. of measurements = 20
"SENSe:MPOWer? 935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20"
'performs the same measurement and in addition returns the results list
immediately after completion of the last measurement.
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
SENSe Subsystem
6.195