R&S ESU Series Operating Manual page 604

Emi test receiver
Table of Contents

Advertisement

R&S ESU
[SENSe<1|2>:]CORRection:TSET[:STATe] ON | OFF
This command switches the selected transducer set on or off.
Example:
"CORR:TSET:SEL 'SET1'"
"CORR:TSET ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R, A
[SENSe<1|2>:]CORRection:TSET:DELete
This command deletes the selected transducer set.
Example:
"CORR:TSET:SEL 'SET1'"
"CORR:TSET:DEL"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value.
[SENSe<1|2>:]CORRection:YIG:TEMPerature:AUTO
This command switches on or off the automatic correction of the YIG filter frequency drift.
When correction is switched on, it is checked once per minute whether the temperature on the YIG
filter has changed by more than 5K relative to the last instance of correction. If this is the case, the
frequency of the YIG filter is – at the end of the next measurement – adjusted as required for the new
temperature. For time-critical applications, the correction function can be switched off after an
operating period of ≥ 30 minutes.
Example:
"CORR:YIG:TEMP OFF"
Characteristics: *RST value: ON
SCPI: device-specific
Mode:
R, A
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
ON | OFF
SENSe Subsystem
6.165