R&S ESU Series Operating Manual page 660

Emi test receiver
Table of Contents

Advertisement

R&S ESU
SOURce<1|2>:FM:DEViation 100Hz to 10MHz
This command defines the maximum frequency deviation at 1 V input voltage at the FM input of the
tracking generator.
The possible value range is 100 Hz to 10 MHz in steps of one decade.
This command is only available in connection with the Tracking Generator Option B9.
Example:
"SOUR:FM:DEV 1MHz"
'Sets the maximum frequency deviation of the tracking generator for screen A to 1
MHz.
Characteristics: *RST value: 100 Hz
SCPI: conform
Mode:
R, A
SOURce<1|2>:FREQuency:OFFSet -150MHz to 150MHz
This command defines a frequency offset of the tracking generator for the indicated measurement
window. Frequency-converting DUTs can be measured with this setting.
The possible value range is -150 MHz to 150 MHz. It should be noted that the terms (start frequency
- tracking frequency offset) and (stop frequency - tracking frequency offset) are both > 1 kHz or both <
-1 kHz.
External I/Q modulation is switched off, if active. This command is only valid in conjunction with the
tracking generator option B9.
Example:
"SOUR:FREQ:OFFS 10MHz"
'Switches on the frequency offset of the tracking generator for screen A to 10 MHz.
Characteristics: *RST value: 0 Hz
SCPI: conform
Mode:
R, A
SOURce<1|2>:POWer[:LEVel][:IMMediate][:AMPLitude] <numeric_value>
This command defines the output level of the tracking generator in the current measurement window.
This command is only valid in conjunction with the tracking generator option B9.
Parameter:
<numeric value>::= -30 dBm to 0 dBm
Example:
"SOUR:POW -20dBm"
'Sets the tracking generator level in screen A to -20 dBm.
Characteristics: *RST value: -20 dBm
SCPI: conform
Mode:
R, A
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
SOURce Subsystem
6.221

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents