INITiate<1|2>:EMITest
In receiver mode, this command starts an automatic sequence consisting of the preliminary
measurement, determination of the peak list and subsequent final measurement.
Example:
"INIT2:EMIT"
'Starts the sequence preliminary measurement, peak search and final
measurement
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R
This command is an event, so it has neither an *RST value nor a query function.
INITiate<1|2>:SPURious
This command starts a new spurious measurement.
Example:
"INIT:CONT OFF"
'switches to single-sweep mode
"INIT:SPUR;*WAI"
'starts the measurement by waiting for the end of the 20 measurements
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A