R&S ESU Series Operating Manual page 589

Emi test receiver
Table of Contents

Advertisement

R&S ESU
File contents
File header
Type;R&S ESU;
Version;3.9x;
Date;02.Feb 2007;
Mode;ANALYZER;SPURIOUS;
Start;9000.000000;Hz
Stop;8000000000.000000;Hz
x-Axis;LIN;
Sweep Count;1;
Range 1:
Start; 9000.000000;Hz
Stop; 150000.000000;Hz
Filter Type;NORMAL;
RBW;10000.000000;Hz
VBW;30000.000000;Hz
Auto Sweep time;ON
Sweep time; 0.145000;s
Detector;RMS;
REF-Level; -10.000000;dBm
Auto RF-Attenuator; OFF;
RF Att;15.000000;dB
Sweep Points;625;
Preamp; 0.000000;dB
Stop after range;OFF;
Transducer;TRD1;
File data section
TRACE 1:
x-Unit;Hz;
y-Unit;dBm;
Values;2500;
9000.000000;-99.619965;
9225.961538;-105.416908;
9451.923077;-100.938057;
9677.884615;-99.483894;
9903.846154;-106.879539;
10129.807692;-108.772316;
6.150
Remote Control – Description of Commands
Explanation
Model
Firmware version
Storage date of data set
Operating mode of the device
Format for Spurious Emissions measurement
Start/stop of the span
Unit: Hz
Linear (LIN) or logarithmic (LOG) (future feature) scaling of the x-
axis
Selected number of sweeps
Loop across all defined sweep ranges (1 to 20)
Range start frequency in Hz
Range stop frequency in Hz
Range filter type: NORMAL, RRC or CFILTER
Resolution bandwidth of the measurement filter
Resolution bandwidth of the video filter
Selected sweep time in the current range
Detector: MAX PEAK, MIN PEAK, RAVERAGE
Setting of the reference level in the current range
Manual (OFF) or automatic (ON) setting of the RF attenuator
Range input attenuation
Number of sweep points in the current range
Range preamplifier, on (20 dB) or off (0 dB)
Stop after range, switched on (ON) or off (OFF) for the current
range
Transducer name (if activated)
repeated for all sweep list ranges
Selected trace
Unit of x values:
Unit of y values:
Number of measurement points
Measurement values:
<x value>; <y value>
Operating Manual 1302.6163.12 - 03
MMEMory Subsystem

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents