[SENSe<1|2>:]LIST:RANGe<1...20>:LIMit:STATe ON | OFF
This command activates/deactivates the limit checking for the spurious emission measurement
EVALUATION
(for details on this measurement see page 220).
The numeric suffix at SENSe<1|2> and the numeric suffix at RANGe1...20> are irrelevant.
Example:
"LIST:RANG5:LIM -40"
'Limit line in range 5 is –40 dB.
"LIST:RANG7:LIM -20"
'Limit line in range 7 is–20 dB.
"LIST:RANG:LIM:STAT ON"
'Activates the limit checking in all ranges.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:POINts <numeric_value>
This command defines the number of sweep points of a range in the spurious measurement.
Example:
"LIST:RANG2:POIN 301"
'Sets 301 sweep points in range 2
Characteristics: *RST value: 625
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:RLEVel <numeric_value>
This command defines the reference level of a range in the spurious measurement.
Example:
"LIST:RANG2:RLEV -30"
'Sets the reference level in range 2 to -30 dBm
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:SWEep:TIME <numeric_value>
This command defines the sweep duration of a range in the spurious measurement.
Example:
"LIST:RANG2:SWE:TIME 2.5MS"
'Sets the sweep time in range 2 to 2.5 ms
Characteristics: *RST value: --
SCPI: device-specific
Mode:
A
Operating Manual 1302.6163.12 - 03