Mode:
A
CALCulate<1|2>:MARKer<1...4>:FUNCtion:FPEaks:X?
This query reads out the list of X values of the maxima found. The number of available values can be
queried with CALC:MARK:FUNC:FPEaks:COUNt?.
With sort mode X, the X values are in increasing order; with sort mode Y the order corresponds to the
decreasing order of the Y values.
Example:
"CALC:MARK:FUNC:FPE:SORT Y"
'sets the sort mode to decreasing y values
"CALC:MARK:FUNC:FPE 3"
'searches the 3 highest maxima for trace 1
"CALC:MARK:FUNC:FPE:COUN?"
'queries the number of maxima found
"CALC:MARK:FPE:FUNC:X?"
'queries the frequencies (span <> 0) or time (span = 0) of the maxima found.
Return value:
"107.5E6,153.8E6,187.9E6"
'frequencies in increasing order
"2.05E-3,2.37E-3, 3.71e-3"
'times in increasing order
Characteristics: *RST value: --
SCPI: device-specific
Mode:
A
CALCulate<1|2>:MARKer<1...4>:FUNCtion:FPEaks:Y?
This query reads out the list of X values of the maxima found. The number of available values can be
queried with CALC:MARK:FUNC:FPEaks:COUNt?.
With sort mode X, the X values are in increasing order; with sort mode Y the order corresponds to the
decreasing order of the Y values.
Example:
"CALC:MARK:FUNC:FPE:SORT Y"
'sets the sort mode to decreasing y values
"CALC:MARK:FUNC:FPE 3"
'searches the 3 highest maxima for trace 1
"CALC:MARK:FUNC:FPE:COUN?"
'queries the number of maxima found
"CALC:MARK:FUNC:FPE:Y?"
'queries the levels of the maxima found.
Return value:
"-37.5,-58.3,-59.6"
'level in decreasing order
Characteristics: *RST value: --
SCPI: device-specific