6.5.4
CALCulate:MARKer Subsystem
The CALCulate:MARKer subsystem checks the marker functions in the instrument.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
In receiver mode, the markers can only be activated after a scan has been carried out.
CALCulate<1|2>:MARKer<1...4>[:STATe] ON | OFF
This command switches on or off the selected marker. If no indication is made, marker 1 is selected
automatically. If marker 2, 3 or 4 is selected and used as a delta marker, it is switched to marker mode.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Example:
"CALC:MARK3 ON"
'Switches marker 3 in screen A on or to marker mode.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:MARKer<1...4>:AOFF
This command switches off all active markers in the indicated measurement window.
All delta markers and active marker/delta marker measurement functions are also switched off.
Example:
"CALC:MARK:AOFF"
'Switches off all markers in screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:TRACe
This command assigns the selected marker (1...4) to the indicated measurement curve. The
corresponding trace must be active, i.e. its status must be different from "BLANK".
If necessary, the corresponding marker is switched on prior to the assignment.
The measurement windows are selected via CALCulate1 (screen A) or 2 (screen B).
Example:
"CALC:MARK3:TRAC 2"
'Assigns marker 3 in screen A to trace 2.
"CALC2:MARK:TRAC 3"
'Assigns marker 1 in screen B to trace 3.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:MARKer<1...4>:X 0 to MAX (frequency | sweep time)
This command positions the selected marker to the indicated :
frequency (span > 0)
6.42