R&S ESU Series Operating Manual page 90

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Meas Time - Entry of measurement time
The measurement time can be set between 100 µs and 100 s separately for each
subrange. In the case of quasipeak weighting, the minimum is 1 ms. The measure-
ment time can be set independently for each scan range.
With option ESU-K53 (Time Domain Scan) the measurement time range depends
on the selected resolution bandwidth.
Remote command:
Auto Ranging - Activates the autorange function
If 0 dB RF attenuation is used with autoranging, care must be taken that the per-
missible signal level at the RF input is not exceeded.
Exceeding this level would damage the input mixer.
The 0 dB attenuation should under no circumstances be used when RFI voltage
measurements are performed with the aid of artificial networks as in this case very
high pulses occur during phase switching.
ON
OFF
Remote command:
RF Atten - Entry of a fixed RF attenuation
The RF attenuation can be set separately for each subrange.
Remote command:
Operating Manual 1302.6163.12 - 03
Resolution
Bandwidth
10 Hz ... 50 Hz
100 Hz ... 500 Hz
1 kHz ... 30 kHz
50 kHz
100 kHz ...120 kHz
200 kHz
300 kHz
500 kHz
1 MHz
R&S ESU automatically sets the input attenuation as a function of
the signal level.
The input attenuation setting of the scan table is used.
Minimum
Measurement Time
10 ms
1 ms
µs
100
µs
100
µs
10
µs
10
µs
10
µs
10
µs
10
SCAN1:TIME 1ms
SCAN1:INP:ATT:AUTO OFF
SCAN1:INP:ATT 10dB
Instrument Functions
Receiver Mode
Maximum
Measurement Time
100 s
100 s
100 s
50 s
30 s
16 s
10 s
6 s
3 s
4.59