R&S ESU Series Operating Manual page 540

Emi test receiver
Table of Contents

Advertisement

R&S ESU
DIAGnostic<1|2>:SERVice:INPut:RECTangle[:STATe] ON | OFF
This command switches the rectangle calibration signal on or off in the selected measurement window
(suffix <1|2>). The setting takes effect only if the RF input is set to the internal reference signal (DIAG:
SERV:INP CAL command).
If the rectangle calibration signal is switched on, the pulsed calibration signal is
switched off.
Example:
"DIAG:SERV:INP CAL"
"DIAG:SERV:INP:RECT ON"
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
R, A
DIAGnostic<1|2>:SERVice:INPut:RECTangle:PRATe 5 kHz | 31.25 kHz | 50 kHz | 250 kHz | 500 kHz
This command sets the frequency of the rectangle calibration signal in the selected measurement
window (suffix <1|2>).
5 kHz | 31.25 kHz | 50 kHz | 250 kHz | 500 kHz can be set.
Example:
"DIAG:SERV:INP:RECT:PRAT 128 MHz
Characteristics: *RST value: 5 kHz
SCPI: device-specific
Mode:
R, A
DIAGnostic<1|2>:SERVice:SFUNction '<string>'
This command activates a service function which can be selected by indicating the five parameters:
function group number, board number, function number, parameter 1 and parameter 2 (see service
manual). The contents of the parameter string is identical to the code to be entered in the data entry
field of manual operation.
The entry of a service function is accepted only if the system password Level 1 or Level 2 has been
entered previously (command: SYSTem:SECurity).
The numeric suffix <1|2> is ignored with this command.
The service functions of the instrument are not identical to those of the FSE family.
That is why the remote command differs in syntax and data format.
Example:
"DIAG:SERV:SFUN '2.0.2.12.1'"
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
DIAGnostic Subsystem
6.101