R&S ESU
NEXT RANGES
Scrolls through the sweep list in forward direction if more than 5 ranges are defined.
The first page shows Range 1 through 5, the second page shows Range 6 to 11 etc.
PREVIOUS
Scrolls through the sweep list in backwards direction if more than 5 ranges are
RANGES
defined. The first page shows Range 1 through 5, the second page shows Range 6
to 11 etc.
REFERENCE
Opens a submenu to define additional parameters for the reference power measure-
RANGE
ment.
POW REF CHAN PEAK
TX BANDWIDTH
RRC FILT (ON OFF)
ALPHA BT
POW REF
Selects the type of power measurement in the reference range.
CHAN PEAK
•
Peak
Measures the highest peak within the reference range.
•
Chan
Measures the channel power within the reference range (integral bandwidth
method).
Remote command:
TX
Defines the bandwidth used for measuring the channel power. Possible values are
BANDWIDTH
from 20 Hz to the span of the reference range.
Remote command:
RRC FILT
Activates or deactivates the RRC filter.
(ON OFF)
Remote command:
ALPHA BT
Sets the rolloff factor of the RRC filter.
Remote command:
START MEAS
Starts the Spectrum Emission Mask measurement. In Single Sweep mode, the
R&S ESU automatically stops the measurement.
Remote command:
STOP MEAS
Stops the Spectrum Emission Maskj measurement.
Remote command:
LIST EVAL
Opens a submenu to edit the list evaluation settings.
LIST EVAL (ON OFF)
MARGIN
4.238
ESP:RTYP PEAK
ESP:BWID 20 HZ
ESP:FILT:RRC:STAT ON
ESP:FILT:RRC:ALPH 0.22
INIT:ESP
ABOR
Operating Manual 1302.6163.12 - 03
Instrument Functions
Analyzer Mode