R&S ESU Series Operating Manual page 269

Emi test receiver
Table of Contents

Advertisement

R&S ESU
NEXT RANGES
Scrolls through the sweep list in forward direction if more than 5 ranges are defined.
The first page shows Range 1 through 5, the second page shows Range 6 to 11 etc.
PREVIOUS
Scrolls through the sweep list in backwards direction if more than 5 ranges are
RANGES
defined. The first page shows Range 1 through 5, the second page shows Range 6
to 11 etc.
REFERENCE
Opens a submenu to define additional parameters for the reference power measure-
RANGE
ment.
POW REF CHAN PEAK
TX BANDWIDTH
RRC FILT (ON OFF)
ALPHA BT
POW REF
Selects the type of power measurement in the reference range.
CHAN PEAK
Peak
Measures the highest peak within the reference range.
Chan
Measures the channel power within the reference range (integral bandwidth
method).
Remote command:
TX
Defines the bandwidth used for measuring the channel power. Possible values are
BANDWIDTH
from 20 Hz to the span of the reference range.
Remote command:
RRC FILT
Activates or deactivates the RRC filter.
(ON OFF)
Remote command:
ALPHA BT
Sets the rolloff factor of the RRC filter.
Remote command:
START MEAS
Starts the Spectrum Emission Mask measurement. In Single Sweep mode, the
R&S ESU automatically stops the measurement.
Remote command:
STOP MEAS
Stops the Spectrum Emission Maskj measurement.
Remote command:
LIST EVAL
Opens a submenu to edit the list evaluation settings.
LIST EVAL (ON OFF)
MARGIN
4.238
ESP:RTYP PEAK
ESP:BWID 20 HZ
ESP:FILT:RRC:STAT ON
ESP:FILT:RRC:ALPH 0.22
INIT:ESP
ABOR
Operating Manual 1302.6163.12 - 03