R&S ESU
Command
Supported subset
VB
VB <numeric_value>
HZ|KHZ|MHZ|GHZ
VB DN
VB UP
VB AUTO
VB?
VBR
1)
VBR <numeric_value>
VBR DN
VBR UP
VBR?
VIEW
VIEW TRA|TRB|TRC
VTL
VTL <numeric_value> DB|DM
VTL DN
VTL UP
VTL?
1) HP 8594E only
2) Command will be accepted without error message, but will be ignored
6.23.3
Special Features of the Syntax Parsing Algorithms for 8566A and
8568A Models
The command syntax is very different for models A and B. Different names are assigned to identical
instrument functions, and the command structure likewise differs considerably between models A and
models B.
The command structure for models A is as follows:
<command> ::=
<command code>[<SPC>][<data>|<step>][<SPC>][<delimiter>][<command code>]...<delimiter>
<data> ::= <value>[<SPC>][<units code>][<SPC>][<delimiter>][<SPC>][<data>]...
<step> ::= UP|DN
where
<command code> = see Table "Supported Commands"
<value> = integer or floating-point numerical value
<units code> = DM | -DM | DB | HZ | KZ | MZ | GZ | MV | UV | SC | MS | US
<delimiter> = <CR> | <LF> | <,> | <;> | <ETX>
<SPC> = 32
10
<ETX> = 3
10
Command sections given in [ ] are optional.
The R&S ESU GPIB hardware differs from that used in the HP analyzers. Therefore, the following con-
straint exists:
<LF>| <EOI> are still used as delimiters since the GPIB hardware is able to identify them. The other
delimiters are identified and evaluated during syntax analysis.
Operating Manual 1302.6163.12 - 03
GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and 8594E
Function
Video
Bandwidth
Video
Bandwidth
Ratio
Video Trigger
Level
Remote Control – Description of Commands
Corresponding
Status
HP-Models
HP 856xE /
available
HP 8566B /
HP 8568B /
HP 8594E
HP 856xE /
available
HP 8566B /
HP 8568B /
HP 8594E
HP 856xE /
available
HP 8566B /
HP 8568B /
HP 8594E
HP 856xE /
available in V3.3x and above
HP 8594E