Change Of Settings Via Markers - Mkr-> Key - R&S ESU Series Operating Manual

Emi test receiver
Table of Contents

Advertisement

R&S ESU
4.8.12
Change of Settings via Markers – MKR-> Key
The MKR-> menu offers functions through which instrument parameters can be
changed with the aid of the currently active marker. The functions can be used on
markers and delta markers.
On opening the menu, the entry for the last active marker is activated; if no marker
was enabled, MARKER 1 is activated and a peak search is performed.
MKR->
SELECT MARKER
PEAK
CENTER = MKR FREQ
REF LEVEL = MKR LVL
NEXT PEAK
NEXT PEAK RIGHT
NEXT PEAK LEFT
SEARCH LIMITS
MKR->TRACE
Side menu
MKR->CF STEPSIZE
MIN
NEXT MIN
NEXT MIN RIGHT
NEXT MIN LEFT
EXCLUDE LO
PEAK EXCURSION
2. side menu
AUTO MAX PEAK
AUTO MIN PEAK
SELECT
The SELECT MARKER softkey activates the numerical selection of the marker in
MARKER
the data entry field. Delta marker 1 is selected by input of ' 0 '.
Remote command:
PEAK
The PEAK softkey sets the active marker/delta marker to the peak of the trace. If no
marker is active when MKR-> menu is called, MARKER 1 is automatically switched
on and the peak search is performed.
Remote command:
4.162
!
LEFT LIMIT /
RIGHT LIMIT
THRESHOLD
SEARCH LIMIT OFF
CALC:MARK1 ON
CALC:MARK1:X <value>
CALC:MARK1:Y?
CALC:MARK:MAX
CALC:DELT:MAX
Instrument Functions
Analyzer Mode
Operating Manual 1302.6163.12 - 03

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents