R&S ESU
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:AVERage ON | OFF
This command switches on or off averaging for the active time domain power measurement in the
indicated window. The function is independent of the marker selection, i.e. the numeric suffix <1...4>
of :MARKer is irrelevant. It is only available in the time domain (span = 0).
Averaging is reset by switching it off and on again.
The number of results required for the calculation of average is defined with [SENSe<1|2>:
]AVERage:COUNt.
It should be noted that synchronization to the end of averaging is only possible in single-sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC2:MARK:FUNC:SUMM:AVER ON"
'Switches on the calculation of average in screen B.
"AVER:COUN 200"
'Sets the measurement counter to 200.
"INIT;*WAI"
'Starts a sweep and waits for the end.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:MODE ABSolute | RELative
This command selects absolute or relative time domain power measurement in the indicated
measurement window. The function is independent of the marker selection, i.e. the numeric suffix
<1...4> of :MARKer is irrelevant. It is only available in the time domain (span = 0).
The reference power for relative measurement is defined with CALCulate:MARKer:FUNCtion:
SUMMary:REFerence:AUTO ONCE. If the reference power is not defined, the value 0 dBm is used.
Example:
"CALC:MARK:FUNC:SUMM:MODE REL"
'Switches the time domain power measurement to relative.
Characteristics: *RST value: ABSolute
SCPI: device-specific
Mode:
A-T
6.84
Remote Control – Description of Commands
CALCulate Subsystem
Operating Manual 1302.6163.12 - 03