R&S ESU
"SENSe:LIST:POWer
performs a measurement sequence with the following settings:
Step
Freq.
Ref Level
[MHz]
1
935.2
-20 dBm
2
935.4
-20 dBm
3
935.6
-20 dBm
"SENSe:LIST:POWer?
performs the same measurement and returns the result list immediately after the last frequency point.
•
The measurement is performed in the time domain and therefore the span is set
to 0 Hz. If the time domain is abandoned, the function is automatically switched
off.
•
The measurement is not compatible with other measurements, especially as far
as marker, adjacent channel power measurement or statistics are concerned.
The corresponding commands thus automatically deactivate the function.
•
The function is only available in remote control. It is deactivated when switching
the instrument back to LOCAL.
Characteristics: *RST value: --
SCPI: device-specific
Mode:
A
6.186
935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0"
RF Att
el Att
Filter type
10 dB
OFF
Normal
10 dB
OFF
Channel
10 dB
OFF
Channel
935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0"
Remote Control – Description of Commands
RBW
VBW
Meas Time
1 MHz
3 MHz
434 us
30 kHz
100 kHz
434 us
30 kHz
100 kHz
434 us
Operating Manual 1302.6163.12 - 03
SENSe Subsystem
TRG Level
(reserved)
0
0
0