R&S ESU Series Operating Manual page 304

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Name/Unit/Interpolation:
Comment:
1.0000000 MHz
1.0000000 GHz
Depending on the softkey selected, either the table with the data of the factor
marked (EDIT TRD FACTOR softkey) or an empty table (NEW FACTOR softkey) is
displayed. This table is empty except for the following entries:
Unit:
Interpolation:
The features of the factor are entered in the header of the table, and the frequency
and the transducer factor are entered in the columns.
Name
Unit
Interpolation
Comment
FREQUENCY
TDF/dB
During editing, a transducer factor remains stored in the background until the factor
edited is saved with the SAVE TRD FACTOR softkey or until the table is closed. A
factor that was edited by mistake can be restored by leaving the entry function.
Name - Entry of name
A maximum of 8 characters is permissible for the name. The characters have to
comply with the convention of DOS file names. If the name exceeds 8 characters,
the name is truncated. The unit automatically stores all transducer factors with the
extension .TDF. If an existing name is changed, the factor stored under the previous
name is retained and will not be overwritten automatically with the new name. The
previous factor can be deleted at a later time, using DELETE. Thus, factors can be
copied.
Remote command:
Operating Manual 1302.6163.12 - 03
EDIT TRANSDUCER FACTOR
FREQUENCY
TDF/dB..
1.000
5.500
dB
LIN for linear frequency scaling
LOG for logarithmic frequency scaling
Entry of name
Selection of unit
Selection of interpolation
Entry of comment
Entry of frequency of reference values
Entry of transducer factor.
Cable
dB
FREQUENCY
CORR:TRAN:SEL <name>
Instrument Functions
Basic Settings
LIN
TDF/dB..
4.273