Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:RMS ON"
'Switches on the function in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:RMS:RES?"
'Outputs the result of screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:RMS:AVERage:RESult?
This command queries the result of the measurement of the averaged RMS value in the selected
measurement window. The query is only possible if averaging has been activated previously using
CALCulate:MARKer:FUNCtion:SUMMary:AVERage ON.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of :MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:RMS ON"
'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:AVER ON"
Switches on the average value calculation in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:RMS:AVER:RES?"
'Outputs the result of screen A.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A-T
This command is only a query and therefore has no *RST value.
6.78
Remote Control – Description of Commands
CALCulate Subsystem
Operating Manual 1302.6163.12 - 03