R&S ESU Series Operating Manual page 627

Emi test receiver
Table of Contents

Advertisement

R&S ESU
[SENSe<1|2>:]LIST:POWer:SET:AVERage:TYPE LINear | LOGarithmic
With this command the average type of the sense list measurement can be set either to linear (default)
or logarithmic. In linear mode voltages are averaged in logarithmic mode levels.
Example:
"LIST:POW:SET:AVER LOG"
'sets log averaging for list measurement.
Characteristics: *RST value: LIN
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:POWer:STATe
This command deactivates the list measurement.
Example:
"SENSe:LIST:POWer:STATe OFF"
Characteristics: *RST value: --
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:BANDwidth[:RESolution] <numeric_value>
This command selects the resolution bandwidth (RBW) of a range in the spurious measurement.
Example:
"LIST:RANG2:BAND 10E3"
'Sets the RBW to 10 KHz.
Characteristics: *RST value: 10 kHz
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:BANDwidth:VIDeo <numeric_value>
This command selects the video bandwidth (VBW) of a range in the spurious measurement.
Example:
"LIST:RANG2:BAND:VIDeo 40E3"
'Sets the VBW to 40 KHz.
Characteristics: *RST value: 30 kHz
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:RANGe<1...20>:BREak ON | OFF
This command determines whether a sweep is stopped in the spurious measurement when a range is
changed.
Example:
"LIST:RANG2:BRE ON"
'Sweep stops if the range is changed from range 2 to 3.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A
6.188
Remote Control – Description of Commands
OFF
SENSe Subsystem
Operating Manual 1302.6163.12 - 03

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents