R&S ESU Series Operating Manual page 834

Emi test receiver
Table of Contents

Advertisement

R&S ESU
9.2 SCPI-Specific Error Messages
Error code
0
Error code
-100
-101
-102
-103
-104
-105
-108
-109
-110
-111
-112
-113
Operating Manual 1302.6163.12 - 03
Error text in the case of queue poll
Error explanation
No error
This message is output if the error queue does not contain any entries.
Error text in the case of queue poll
Error explanation
Command Error
The command is faulty or invalid.
Invalid Character
The command contains an invalid sign.
Example: A header contains an ampersand, "SENSe&".
Syntax error
The command is invalid.
Example: The command contains block data the instrument does not accept.
Invalid separator
The command contains an invalid sign instead of a separator.
Example: A semicolon is missing after the command.
Data type error
The command contains an invalid value indication.
Example: ON is indicated instead of a numeric value for frequency setting.
GET not allowed
A Group Execute Trigger (GET) is within a command line.
Parameter not allowed
The command contains too many parameters.
Example: Command SENSe:FREQuency:CENTer permits only one frequency
indication.
Missing parameter
The command contains too few parameters.
Example: The command SENSe:FREQuency:CENTer requires a frequency
indication.
Command header error
The header of the command is faulty.
Header separator error
The header contains an invalid separator.
Example: the header is not followed by a "White Space", "*ESE255"
Program mnemonic too long
The header contains more than 12 characters.
Undefined header
The header is not defined for the instrument.
Example: *XYZ is undefined for every instrument.
Error Messages
SCPI-Specific Error Messages
9.3