R&S ESU Series Operating Manual page 308

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Remote command:
Break - Activation of query when changing a subrange
The sweep can be stopped if the user changes the subrange and selects a new sub-
range of the transducer. A message informs the user that the limit has been
attained. He can continue the sweep or switch off the transducer.
The interruption is activated by setting Break to ON. Selection is by the ENTER key
which toggles between ON and OFF (toggle function).
Remote command:
Start - Entry of start frequency of subrange
Stop - Entry of stop frequency of subrange
The individual subranges have to be linked without a gap. That is why the start fre-
quency is already defined from the second subrange (= stop frequency of previous
range).
Remote command:
INS BEFORE
The INS BEFORE RANGE softkey copies the active column and inserts it to the
RANGE
right. The frequency limits can be edited.
Remote command:
INS AFTER
The INS AFTER RANGE softkey copies the active column and inserts it to the left.
RANGE
The frequency limits can be edited.
Remote command:
DELETE
The DELETE RANGE softkey deletes the marked subrange. The following sub-
RANGE
ranges move left.
Remote command:
RANGES 1-5/6-
The RANGES 1-5/6-10 softkey switches between the display of ranges 1 to five and
10
ranges 6 to 10.
Remote command:
Operating Manual 1302.6163.12 - 03
CORR:TSET:UNIT <string>
CORR:TSET:BRE ON|OFF
CORR:TSET:RANG<1.10>
<freq>,<freq>,<name>..
--
--
--
--
Instrument Functions
Basic Settings
4.277

Hide quick links: