R&S ESU Series Operating Manual page 657

Emi test receiver
Table of Contents

Advertisement

R&S ESU
Example:
"SWE:IF:SHIF AUTO"
'Sets the IF shift to mode AUTO.
"SWE:IF:SHIF:FREQ <nummeric value>"
'Specifies the signal frequency for IF SHIFT AUTO.
Characteristics: *RST value: OFF
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]SWEep:IF:SHIFt:FREQuency <numeric value>
This command defines the signal frequency to automatically select the suitable 1
IF SHIFT AUTO (SENS:SWE:IF:SHIFT:MODE AUTO).
Example:
"SWE:IF:MODE AUTO"
'activates IF SHIFT AUTO.
"SWE:IF:SHIF:FREQ 2GHz"
'carrier signal at 2 GHz.
Characteristics: *RST value: 1 GHz
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]SWEep:MODE AUTO | LIST | ESPectrum
This command controls the sweep continuation.
AUTO
free running sweep - the spurious list is activated
ESPectrum
activates the Spectrum Emission Mask measurement
LIST
Activates the spurious emissions measurement. The sweep list is
defined using the SENS:LIST:RANGE commands.
Example:
"SWE:MODE LIST"
Activates the spurious emissions measurement.
Characteristics: *RST value: AUTO
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]SWEep:POINts 155, 313, 625, 1251, 1999, 2501, 5001, 10001, 20001, 30001
This command defines the number of measurement points for one sweep run.
Example:
"SWE:POIN 313"
Characteristics: *RST value: 625
SCPI: conform
Mode:
A
6.218
Remote Control – Description of Commands
st
Operating Manual 1302.6163.12 - 03
SENSe Subsystem
IF shifting for mode