R&S ESU Series Operating Manual page 500

Emi test receiver
Table of Contents

Advertisement

R&S ESU
CALCulate<1|2>:MARKer<1...4>:FUNCtion:CSTep
This command sets the step width of the receiver or center frequency in the selected measurement
window to the X value of the current marker.
If marker 2, 3 or 4 is selected and used as delta marker, it is switched to the marker mode.
Example:
"CALC2:MARK2:FUNC:CST"
'Sets the receiver frequency to the same value as the frequency of marker 2.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
R, A-F
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:REFerence
This command sets the reference level to the power measured by the indicated marker. If the selected
marker was used as delta marker, it is switched to marker mode.
Example:
"CALC:MARK2:FUNC:REF"
'Sets the reference level of screen A to the level of marker 2.
Characteristics: *RST value: -
SCPI: device-specific
Mode:
A
This command is an event and therefore has no *RST value and no query.
Operating Manual 1302.6163.12 - 03
Remote Control – Description of Commands
CALCulate Subsystem
6.61

Hide quick links: